AC power supply PCR-LE series

Advanced


Fixing the Internal Vcc

To minimize loss in the linear amplifier section, the PCR-LE Series automatically adjusts the linear amplifier supply voltage (Vcc) to a level that is suitable for the output voltage. Because the response speed of the preregulator (DC power supply) that generates Vcc is slower than that of the linear amplifier, the Vcc voltage is set internally (only when the output voltage is being raised) several hundreds of milliseconds before the voltage output is changed. There will be a time lag before the output voltage is actually changed even when you frequently change the output voltage externally.

111118.jpg CAUTION

When the internal Vcc is set to FIXED while the PCR-LE Series operates in the DC mode, the operating range will change. Operating the PCR-LE Series outside the operating range will cause a malfunction.

 

You can fix the Vcc voltage of the PCR-LE Series. This is useful when you want to prioritize the output voltage response over the product’s efficiency.

Fix the PCR-LE Series’ Vcc voltage according to the peak voltage in the output voltage change tests.

When you fix the Vcc, the internal (power) loss increases, so the protection functions may be activated depending on the output voltage (the duration that the voltage is generated for), load conditions, and ambient temperature. Be sure to carry out tests (simulations) in advance, and check how the PCR-LE Series operates.

If the Vcc is fixed and the maximum used output voltage is set lower than the peak output voltage, the maximum used output voltage setting is changed to follow the peak output voltage to prevent a malfunction.

Press OTHERS (SHIFT+MEMORY), 1/3 (F6), and then VCC (F1) to set the internal Vcc.

35104.jpg 

Item

Title

Description

FIXED

VCCVolt

Sets the maximum output voltage that is used (0 Vpk to 431.0 Vpk).

The Vcc voltage is fixed.

AUTO

VCCVolt

The Vcc voltage is set automatically (it is not fixed).

Operating range of DC mode

When the internal Vcc is set to FIXED while the PCR-LE Series operates in the DC mode, the operating range will change. Do not operate the product outside the operating range.

The graph below shows the operating ranges when the maximum used output voltage setting in the L range is fixed to 50 V, 100 V, 150V, or 215.5 V, and that in the H range is fixed to 100 V, 200 V, 300 V, or 431 V.

Operating the product outside the operating range A will cause a malfunction.

Operating range when Vcc is set to 50 V (L range) or 100 V (H range).

111162.jpg 

Operating range when Vcc is set to 100 V (L range) or 200 V (H range).

111185.jpg 

 

Operating range when Vcc is set to 150 V (L range) or 300 V (H range).

111194.jpg 

Operating range when Vcc is set to 215.5 V (L range) or 431 V (H range).

111208.jpg 

 