AC power supply PCR-LE series
Advanced
Setting Soft Starts (The voltage rise time)
To prevent the output from being turned off (the alarm from being activated) and the voltage from dropping due to the load device’s inrush current that exceeds the rated capacity of the PCR-LE Series, you can control the inrush current by having the output voltage rise gradually when the output is turned on.
Press OTHERS (SHIFT+MEMORY) and then RISETIM (F1) to set the soft start.
Item |
Title |
Description |
Condition in which the function key cannot be used |
ON |
Rise Time |
Sets the rise time (0.1 s to 3.0 s). The soft start function is turned on. |
Action to perform when the current limit is exceeded: DISABLE Compensation function: soft sensing or regulation adjustment |
OFF |
The soft start function is turned off. |
None |