AC power supply PCR-LE series
Advanced
The PCR-LE Series’ output impedance (output resistance) is approximately 0 Ω. Commercial power supplies have an impedance (resistance) of several milliohms to several ohms. You can set the PCR-LE Series’ output impedance. This enables you to simulate the same environment as that which is provided by commercial power supplies.
The impedance that is required for IEC standard tests has a prescribed inductance component, but the output impedance of the PCR-LE Series consists entirely of resistance. You can use the PCR-LESeries to perform rough harmonic current analyses. If you require accurate data, use the Line Impedance Network (LIN Series).
Press OTHERS (SHIFT+MEMORY) and then OUT IMP (F3) to set the output impedance.
Item |
Title |
Description |
Condition in which the function key cannot be used |
ON |
OutZ |
Sets the output impedance as a percentage (1 % to 100 %). The output impedance will be set. |
Compensation function: any setting but “OFF” |
OFF |
The output impedance will not be set. |
None |
The values in the following table are assigned to the output resistance percentages (1% to 100%).
|
PCR |
PCR |
PCR |
PCR |
PCR |
PCR |
PCR |
L range |
0 Ω to |
0 Ω to |
0 Ω to |
0 Ω to |
0 Ω to |
0 Ω to |
0 Ω to |
H range |
0 Ω to |
0 Ω to |
0 Ω to |
0 Ω to |
0 Ω to |
0 Ω to 1.333 Ω |
0 Ω to |
AC power supply PCR-LE series