OPERation:PROTecting status register
The PROTecting status register indicates information when a protection or limit function is activated.
Table 11 PROTecting status register (STATus:OPERation:PROTecting)
Bit | Bit Weight | Bit Name | Description |
0 | 1 | OV (Over Voltage Protection) | Overvoltage protection activated |
1 | 2 | UV (Under Voltage protection) | Undervoltage protection activated |
2 | 4 | OC (Over Current protection) | Overcurrent protection activated |
3 | 8 | OP (Over Power protection) | Overpower protection activated |
4 | 16 | OT (Over Temperature protection) | Overtemperature protection activated |
5 | 32 | Not Used | Not used |
6 | 64 | EXT Problem (External Problem) |
Alarm input detected (PLZ-4W) |
7 | 128 | REV (Reverse Voltage) | Reverse connection detected |
8 | 256 | Not Used | Not used |
9 | 512 | Not Used | Not used |
10 | 1024 | Not Used | Not used |
11 | 2048 | Not Used | Not used |
12 | 4096 | Not Used | Not used |
13 | 8192 | Not Used | Not used |
14 | 16384 | USR (User ALARM Input) | Alarm input detected (KFM2150) |
15 | 32768 | Not Used | Always zero |
Queries the event of the PROTecting status register.
A query clears the contents of the register.
Command
STATus:OPERation:PROTecting[:EVENt]?
Response
Returns the event of the PROTecting status register in the NR1 form.
Queries the condition of the PROTecting status register.
A query does not clear the contents of the register.
Command
STATus:OPERation:PROTecting:CONDtion?
Response
Returns the condition of the PROTecting status register in the NR1 form.
Sets the enable register of the PROTecting status register.
Command
STATus:OPERation:PROTecting:ENABle <NR1>
STATus:OPERation:PROTecting:ENABle?
Parameter
Value: 0 to 32767
Response
Returns the enable register setting of the PROTecting status register in the NR1 form.
Sets the positive transition of the PROTecting status register.
Command
STATus:OPERation:PROTecting:PTRansition <NR1>
STATus:OPERation:PROTecting:PTRansition?
Parameter
Value: 0 to 32767
Response
Returns the positive transition of the PROTecting status register in the NR1 form.
Sets the negative transition of the PROTecting status register.
Command
STATus:OPERation:PROTecting:NTRansition <NR1>
STATus:OPERation:PROTecting:NTRansition?
Parameter
Value: 0 to 32767
Response
Returns the negative transition of the PROTecting status register in the NR1 form.